No lights showing during the emWave charging process.
Insert the battery charger into a wall socket. Plug the battery charger into the emWave charging jack. Sometimes, after plugging the emWave into the battery charger, the unit may remain dark (no lights showing at all). Wait for 15 - 20 minutes, sometime during this time the charging indicator light will go on. Continue to charge the unit for 2 hours past this point and the unit will be fully charged. Occasionally, when you first plug the unit into the battery charger, you will only see a quick flash of 2 solid lights at the bottom of the breath pacer then the unit goes dark completely. This is indicative of an incomplete connection in the wall socket. Try using a different wall socket. You should leave the emWave unit on the battery charger for 2 hrs. at which point the unit should be fully charged and remove it from the battery charger
|